Aalborg PWE Digital User Manual

Page 79

Advertising
background image

74

INDEX

NAME

DATA TYPE

NOTES

33

TotalFlowStart

float

Start Main Totalizer at flow [%FS] 0 - Disabled

34

TotalVolStop

float

Main Totalizer Action Limit Volume [%*s]
0- Disabled

35

TotalConfLock

uint

Key Pad Totalizer reset access Lock [0 – Disabled, 1
– Enabled]

36

UDUnitKfactor

float

K-Factor for User Defined Units of Measure

37

UDUnitTimeBase

int

K-Factor = UDUnit/(L/min)

38

UDUnitDensity

uint

User Defined Unit Time Base [1, 60, 3600 sec]

39

FoutScaleV

float

User Defined Unit Density Flag
[0-not used, 1 - used]

40

FoutOffsetV

float

Flow Analog 0-5 Vdc Out Scale

41

FoutScalet_mA

float

Flow Analog 0-5 Vdc Out Offset

42

FoutOffset_mA

float

Flow Analog 4-20 mA Out Scale

43

ToutScaleV

float

Temperature Analog 0-5 Vdc Out Scale**

44

ToutOffsetV

float

Temperature Analog 0-5 Vdc Out Offset**

45

ToutScalet_mA

float

Temperature Analog 4-20 mA Out Scale**

46

ToutOffset_mA

float

Temperature Analog 4-20 mA Out Offset**

47

FlowMeasureMode

uint

0-Pulse width measure, 1 - Number of Pulses per
measure interval

48

OptOut1_Config

uint

Optical Output #1 Configuration (function) [0-10]

49

OptOut2_Config

uint

Optical Output #2 Configuration (function) [0-10]

50

RTD_LinearMode

uint

RTD Linearizer (0-Disabled, 1-Enabled)**

51

AlarmLatch

uint

Alarm Latch settings [0-3]

52

PTotalMode

uint

Pilot Totalizer mode (0-Disabled, 1-Enabled)

53

Reserved

uint

Reserved

54

PTotalFlowStart

float

Start Pilot Totalizer at flow [%FS] 0 - Disabled

55

PTotalVolStop

float

Pilot Totalizer Action Limit Volume [%*s]
0-Disabled

56

MeterFSRange

float

Meter Full Scale range in L/min

57

LowFlowCutOff

float

Low Flow cut off. Must be between 0 and 10.0
%F.S.

58

Damping

uint

Flow Reading Damping 1-99 seconds

59

Density

float

Fluid Density g/cm3 [0.01 - 5.00000 g/cm3]

60

PulseMeasInt

uint

Flow Pulse Measure Interval in mS [500-60000]

61

MeterCalFactor

uint

Calibration Factor:Number of Pulses per Gallon

62

FlowTbl[0].FlowPFS

float

Flow Linearizer Index 0 PFS (must be 0.0)

63

FlowTbl[0].LinCounts

uint

Flow Linearizer Index 0 Counts (must be 0)

64

FlowTbl[1].FlowPFS

float

Flow Linearizer Index 1 PFS [0.0 – 1.0]

65

FlowTbl[1].LinCounts

uint

Flow Linearizer Index 1 Counts

66

FlowTbl[2].FlowPFS

float

Flow Linearizer Index 2 PFS [0.0 – 1.0]

67

FlowTbl[2].LinCounts

uint

Flow Linearizer Index 2 Counts