5.2.2 Flow 0-5 Vdc analog output calibration

1. Install jumpers J3A, J3B and J3C on the PC board for 0-5 Vdc output

(see Table 5-1).

2. Connect a certified high sensitivity multi meter set for the voltage

measurement to the pins M (+) and K (-) of the PWE 12 Pin "M16" connector.

3. Write 4000 counts to the DAC_A channel: !11,WRITE,0,4000[CR]
4. Read voltage with the meter and calculate FOutScaleV value:

5. Save FOutScaleV in to the EEPROM: !11,MW,39,X[CR]

Where: X – the calculated FoutScaleV value.

5.2.3 Flow 4-20 mA analog output calibration

1. Install jumpers J3A, J3B and J3C on the PC board for 4-20 mA output

output (see Table 5-1).

2. Connect a certified high sensitivity multi meter set for the current

measurement to pins M (+) and K (-) of the PWE 12 Pin "M16" connector.

3. Write 4000 counts to the DAC_A channel: !11,WRITE,0,4000[CR]
4. Read current with the meter and calculate FoutScale_mA value:

5. Write zero counts to the DAC_A channel: !11,WRITE,0,0CR]
6. Read offset current with the meter and calculate FoutOffset_mA value:

7. Save FoutScale_mA in to the EEPROM: !11,MW,41,Y[CR]

Save FoutOffset_mA in to the EEPROM: !11,MW,42,Z[CR]

Where: Y – the calculated FoutScale_mA value.

Z – the calculated FoutOffset_mA value.

56

FoutScaleV=

20000

Reading[V]

FoutScale_mA =

4000

Reading[mA]

FoutOffset_mA=-FOutScale_mA*Offset_Reading[mA]