Section 3. CR9000X Measurement Details

The integration and Sinc filter order that a given CR9058E is
using can be seen through RTDaq's terminal mode window
(DataLogger/ Terminal Emulator) or any other terminal
emulator. Click on "Open Terminal", then hit "Enter" several
times until the CR9000> prompt is returned. Type in "4" and
enter. The CR9058Es' slot numbers, integration times, and Sinc
filters will be returned.

TIP

FIGURE 3.2.2-1. Depiction of a 5th order Sinc filter

As shown, the number of samples required to feed a fifth order Sinc filter with
an integration of 288 uSec is 11. The number of samples required for filter
orders of 2 and above can be calculated using equation 3.2.2.

Eq. 3.2.2

(

)

1

)

1

96

(

+

=

IntegTime

r

FilterOrde

mples

NumberofSa

The CR9058 firmware limits the number of samples to 680 in order to reduce
the amount of time required to compute the weighting coefficients. If the
calculated NumberofSamples is greater than 680, then the Filter order is
incrementally reduced until either the Number of Samples is less than 680, or
the Filter Order is 1. A filter order of 1, simple averaging, does not require
storing multiple values.

Solving equation 3.2.2 for the maximum integration time based on the filter
order results in:

Eq. 3.2.3

96

)

1

(

+

=

r

FilterOrde

fSamples

MaxNumbero

Integer

me

MaxIntegTi

Or

96

)

1

680

(

+

=

r

FilterOrde

Integer

me

MaxIntegTi

3-25