Campbell Scientific CR9000X Measurement and Control System User Manual

Page 367

Advertising
background image

Section 9. Program Control Instructions

ZERO CALIBRATION STEPS (Function = 10)

1. If the Reps and Index parameters are constants, go to Step 2.
If

the

Reps and Index parameters are variables then either:

A.

Individual Sensor Cal: Set the Reps parameter to 1 and select the

individual sensor to be zeroed by setting the Index parameter or;

B.

Complete Array Cal: Set the Index parameter to 1 and the Reps

parameter to the number of elements in the MeasVar variable array.

This will zero all of the elements of the array together.

2. Change the Mode value to 1 while the sensor(s) are at their Zero state. The

current mV per volt output from the Bridge measurement will be used for the

Zero argument of the StrainCalc instruction. After the calibration is

complete, the logger will change the Mode value to 6.

SHUNT CALIBRATION STEPS (Function = 13, 33, or 34)

1. Set the Index parameter, if a variable, to point to the element of the

MeasureVar array on which to perform the calibration. Make sure that the

Reps parameter's value is set to 1.

2. Change the value of the correct element of the KnownR array to the

resistance, in ohms, of the strain gauge that will be shunted. At the un-

shunted condition, change the Mode value to 1. The logger will record this

first point's micro-strain value and then change the Mode value to 3.

3. While the Mode value is 3, apply the shunt resistor across one of the arms of

the wheatstone bridge.

Load the shunt resistance value (ohms) into the KnownR parameter as a

positive number if shunting across:

The arm that holds strain gauge for Function 13

The arm that holds gauge that is parallel to

for Function 33

An arm that holds gauge that is parallel to

for Function 43

Load the shunt resistance value (ohms) into the KnownR parameter as a

negative number if shunting across:

The arm that holds completion resistor for Function 13

The arm that holds gauge that is parallel to

ε

− for Function 33

An arm that holds gauge that is parallel to

ε

− for Function 43

Using the correct sign notation on the input resistance of the shunt insures

that the correct polarity is returned (positive strain for tension and negative

for compression). A gauge parallel to

is a gauge that experiences tension

when the element that it is mounted on experiences positive strain. A gauge

parallel to

ε

− is a gauge that experiences compression when the element that

it is mounted on experiences positive strain. See the Function parameter for

code definitions.

When performing a shunt calibration on a bridge with 1 active element

(Function 13: Quarter Bridge Strain), if possible, it is preferable to remotely

shunt across the arm containing the strain gauge as shown with shunt resistor

R1, used with one of our TIMs, in Figure 1A. With this setup, the shunt

resistor value would be entered as a positive value.

If it is not possible to shunt across the gauge, due to accessibility problems, it

is possible to shunt across the bridge arm containing the dummy resistor right

at the datalogger. This shunt setup is depicted with shunt resistor R2 in

Figure 1B. The shunt resistor value would be entered as a negative value.

When performing a shunt calibration on a bridge with 2 active elements

(Function 23: Half Bridge Strain), or with 4 active elements (Function 33:

Full Bridge Strain), the shunt must be done directly across one of the active

gauges.

4. After the shunt is in place, with the shunt ohm resistance value loaded in

KnownR, change the mode value to 4. The datalogger will do the required

calculations, adjust the gauge factor, and change the Mode value to 6.

9-37

Advertising