21 1 f v v – Campbell Scientific CR9000X Measurement and Control System User Manual

Page 127

Advertising
background image

Section 3. CR9000X Measurement Details

circuit, a larger input voltage transition is required for higher frequencies. The
transition required for the input of the Schmitt trigger can be viewed as 2.5
volts

±

1 volt (from below 1.5 volt to above 3.5 volt). The equation to

calculate the amount that the signal is attenuated by the front end filter is:

( )

(

)

(

)

2

2

1

1

f

V

V

In

Out

πτ

+

=

V

Out

is the voltage level leaving the filter (level into the Schmitt circuit) when

V

In

is the input voltage. V

Out

must be at minimum 1 volt for the Schmitt

circuit to trigger an output.

Chart 3.4.4-1 Required Transition Voltage for High Frequency Pulse

0.6

0.8

1

1.2

1.4

1.6

0

100000

200000

300000

400000

500000

600000

700000

800000

900000 1000000 1100000

Signal Frequency ( Hz)

±V Transistion (Centered at 2.5 Volt)

RequiredTransition Voltage

Chart 3.4.4-1 plots the trace for the minimum transition voltage about 2.5 volts
against the input signal frequency. To demonstrate how to use this plot, for a
input frequency of 1 MHz, the voltage signal, centered about 2.5 volts, must
have a transition of

±

1.6 volts in order to trigger the Schmitt circuit. In other

words, the signal must rise from below 0.9 volts (2.5 volts minus 1.6 volts) to
above 4.1 volts (2.5 volts plus 1.6 volts) for a pulse to be counted.

The input voltage range for the Pulse channels is

±

20 V.

Voltages outside of this range can damage the logger.

NOTE

I/O 1 – 16

When using the CR9071E's I/O ports for pulse timing (TimerIO instruction),
the positive threshold voltage is 3.5 V and the negative threshold voltage is 1
V. The maximum input voltage allowed is 5.5 volts and the minimum voltage
allowed is -0.5 V. Voltages outside of this range can damage the CR9071E.

3-39

Advertising