Brhalf, brhalf3w, brhalf4w parameters – Campbell Scientific CR9000X Measurement and Control System User Manual

Page 236

Advertising
background image

Section 7. Measurement Instructions

7-12

Parameter

Enter

BRHALF, BRHALF3W, BRHALF4W PARAMETERS

Dest

Var. or Array

The Variable in which to store the results of the instruction. When Reps are used the results are stored in an
array with the variable name. An array must be dimensioned to have elements for all the Reps.

Reps

Constant

The number of repetitions for the measurement.

Range

The voltage range for the measurement.

Constant

± 5 Volt Analog Input Module

± 50 Volt Analog Input Module

Alpha
Code

Numeric
Code

Voltage
Range

Alpha
Code

Numeric
Code

Voltage
Range

mV5000

0

± 5000 mV

V50

6

± 50 V

mV1000

1

± 1000 mV

V10

7

± 10 V

mV200

4

± 200 mV

V2

10

± 2 V

mV50

5

± 50 mV

mV500

11

± 500 mV

ASlot

Constant

The number of the slot that holds the Analog Input Module to be used for the measurement.

SEChan

Constant

The single-ended channel number on which to make the first measurement. When Reps are used, subsequent
measurements will be made on sequential single-ended channels.
Burst Option

: Enter a negative number to dwell on the specified channel for the # of measurements specified

by the Reps parameter (except for CR9058E). When using burst option, the MesPEx parameter must be set to
the same value as the Reps parameter and a CAO should be used for excitation.

ExSlot

Constant

The slot that holds the Excitation Module for the measurement.

ExChan

Enter the excitation channel number to excite the first measurement.

Constant

Channels Result

1 - 6

Continuous analog output channels, will remain at the excitation voltage set by the
instruction unless a subsequent instruction changes their voltage

7 - 16

Switched excitation channels, are switched to the excitation voltage for the
measurement and switched off between measurements.

MesPEx

Constant

The number of sensors to excite with the same excitation channel before automatically advancing to the next
excitation channel. To excite all sensors with one excitation channel, MesPEx should equal Reps.

ExmV

Constant

The excitation voltage in millivolts. Allowable range ± 5000 mV. RevEx may be used to excite with both a
positive and negative polarity to cancel offset voltages.

RevEx

Option to reverse excitation to cancel offsets. See Section 3.1.1.1 Reversing Excitation or Differential Input.

Constant

Value Result

0

Excite only with the excitation voltage entered

1

A second measurement is made with the voltage polarity reversed.

RevDiff

Option to reverse inputs to cancel offsets. See Section 3.1.1.1 Reversing Excitation or Differential Input.

Constant

Value Result

0

Signal is measured with the high side referenced to the low

1

A second measurement is made after reversing the inputs

SettlingTime

Constant

The time in microseconds to delay between setting up a measurement (switching to the channel, setting the
excitation) and making the measurement (10 microsecond resolution).See Section 3.1.3 Signal Settling Time


Entry


Voltage Range


Delay

CR9055
Voltage Range


Delay

0

± 50 mV

20 µS (default)

± 0.5V

40 µS (default)

0

± 200 mV

20 µS (default)

± 2 V

40 µS (default)

0

± 1000 mV

10 µS (default)

± 10 V

30 µS (default)

0

± 5000 mV

10 µS (default)

± 50 V

30 µS (default)

> 0

all

Truncate to closest 10 µS

all

Truncate to closest 10 µS

Integ

Constant

The integration time in microseconds for each of the channels measured (10 microseconds resolution).
See Section 3.1.1.3 Integration for more information on Integration.

Mult, Offset

Constant, Var.,
Array, or
Expression

A multiplier and offset by which to scale the raw results of the measurement.
The raw result with a multiplier of 1 and an offset of 0 is V

Out

/V

EX

for the BrHalf instruction, and is

V

RS

/V

RF

for the BRHalf3W and BrHalf4W bridge configurations.

See text above to convert this over to

sensor resistance value.