Tcse parameters – Campbell Scientific CR9000X Measurement and Control System User Manual

Page 232

Advertising
background image

Section 7. Measurement Instructions

7-8

Parameter

Enter

TCSE PARAMETERS

Dest

Var. or Array

The Variable in which to store the results of the instruction. When Reps are used the results are stored in
an array with the variable name. An array must be dimensioned to have elements for all the Reps.

Reps

Constant

The number of repetitions for the measurement.

*Range

The voltage range for the measurement. TCSE raw output is in Celsius.

Constant

± 5 Volt Analog Input Module

± 50 Volt Analog Input Module

Alpha
Code

Numeric
Code

Voltage
Range

Alpha
Code

Numeric
Code

Voltage
Range

mV5000

0

± 5000 mV

V50

6

± 50 V

mV1000

1

± 1000 mV

V10

7

± 10 V

mV200

4

± 200 mV

V2

10

± 2 V

mV50

5

± 50 mV

mV500

11

± 500 mV

Aslot

Constant

The number of the slot that holds the Analog Input Module to be used for the measurement.

SEChan

Constant

The single-ended channel number on which to make the first measurement. When Reps are used,
subsequent measurements will be made on sequential single-ended channels. Enter a negative number to
dwell on that channel for the number of measurements specified by the Reps parameter (except for
CR9058E).

TCType

The code for the thermocouple type.

Constant

Alpha Code

Numeric Code

Thermocouple Type

TypeT

0 Copper

Constantan

TypeE

1 Chromel

Constantan

TypeK

2 Chromel

Alumel

TypeJ

3 Iron

Constantan

TypeB

4 Platinum

Rhodium

TypeR

5 Platinum

Rhodium

TypeS

6 Platinum

Rhodium

TypeN

7 Nicrosil-Nisil

(NiCRSi-NiSiMg)

TRef

Variable

The name of the variable that is the reference temperature for the thermocouple measurements.

SettlingTime

Constant

The time in microseconds to delay between setting up a measurement (switching to the channel, setting
the excitation) and making the measurement (10 microsecond resolution)
See Section 3.1.3 Signal Settling Time.


Entry


Voltage Range


Delay

CR9055
Voltage Range


Delay

0

± 50 mV

20 µS (default)

± 0.5V

40 µS (default)

0

± 200 mV

20 µS (default)

± 2 V

40 µS (default)

0

± 1000 mV

10 µS (default)

± 10 V

30 µS (default)

0

± 5000 mV

10 µS (default)

± 50 V

30 µS (default)

> 0

all

Truncate to closest 10 µS

all

Truncate to closest 10 µS

Integ

Constant

The integration time in microseconds for each of the channels measured (10 microseconds resolution).
See Section 3.1.1.3 Integration for more information on Integration.

Mult, Offset

Constant, Var,
Array, or
Expression

A multiplier and offset by which to scale the raw results of the measurement. The raw result of the
TCDiff instruction is in degrees C. A multiplier of 1.8 and an offset of 32 will convert the temperature
to degrees F.

*Range: See notes in TCDiff section.

See Section 3.1.4 Thermocouple Measurements for an in-depth study of TC
measurements and an error analysis for them.

See Section 3.1.2.1 Single Ended Voltage Measurements for in-depth
coverage of the Single Ended Measurement process.

Advertising