Section 7. Measurement Instructions

7-45

Parameter

Enter

VOLTFILT PARAMETERS

Dest

Variable,Array

The Variable to store the results of the instruction. When Reps are used the results are stored in an
array with the variable name. An array must be dimensioned to have elements for all Reps.

Reps

Constant

The number of times to repeat the measurement on subsequent CR9052 channels.

.

Range

The voltage range for the measurement. The CR9052 normally replaces out-of-range

measurements with not-a-number (NaN) which is displayed in RTDaq's real time windows as

Range?

. Users may choose to have out-of-range measurements to be replaced by the analog-to-

digital converter saturation value with a special code in FiltOption.

Constant

Alpha
Code

Numeric
Code

Voltage
Range

Module Excitation
Board Supported

mV5000

0

± 5000 mV

CR9052DC, CR9052IEPE

mV1000

1

± 1000 mV

CR9052DC, CR9052IEPE

mV200

4

± 200 mV

CR9052DC

mV50

5

± 50 mV

CR9052DC

mV20

6

± 20 mV

CR9052DC

Fslot

Constant

The number of the slot that holds the CR9052 Module to be used for the measurement.

Chan

Constant

The CR9052 channel number on which to make the first measurement. When Reps are used,
subsequent measurements will be automatically made on the following differential channels.

FiltOption
Constant

The sample ratio for the measurement (how many measurements are made within one cycle of the
highest frequency in the pass band). The sample ratio determines the top of the pass-band (F

PASS

)

and the beginning of the stop-band (F

STOP

) of the anti-aliasing low-pass filter relative the sample

rate (F

SAMPLE

). The sample rate is the inverse of the scan interval in the CRBASIC program.

FiltOption

must be the same for all channels of a single CR9052 Filter Module. Out-of-range

measurements may be replaced by the analog-to-digital converter saturation value by adding 1000
to the FiltOption codes shown below.

NumericCode Sampling

Ratio

F

PASS

F

STOP

2 2.5

F

SAMPLE

/2.5

F

SAMPLE

/2.01

5 5

F

SAMPLE

/5

F

SAMPLE

/3.37

10 10

F

SAMPLE

/10

F

SAMPLE

/5.08

20 20

F

SAMPLE

/20

F

SAMPLE

/6.81

1* 2.155

23.2 Khz

26.8 kHz

*

Option 1 has no additional filtering beyond the CR9052DC analog front-end and the sigma-delta

A/D converter, thus freeing the CR9052DC on-board digital signal processor for additional
processing. F

SAMPLE

must be 50 kHz to use this filter option.

Excitation

Constant

The continuous, dc, output for the excitation channel(s). If Reps is greater than one, then the same
excitation will be output on sequential excitation channels.

Numeric Code

Alpha Code

Output Level

IEPE Freq. Response

CR9052IEPE

900 (905)

None

None

τ

= 0.5 Sec (

τ

= 5.0 Sec)

CR9052IEPE

605

None

Constant 6 mA

0.3 Hz to 20 kHz (

τ

= 5.0 Sec)

CR9052IEPE

405

None

Constant 4 mA

0.3 Hz to 20 kHz (

τ

= 5.0 Sec)

CR9052IEPE

205

None

Constant 2 mA

0.3 Hz to 20 kHz (

τ

= 5.0 Sec)

CR9052IEPE

600

None

Constant 6 mA

3.0 Hz to 20 kHz (

τ

= 0.5 Sec)

CR9052IEPE

400

None

Constant 4 mA

3.0 Hz to 20 kHz (

τ

= 0.5 Sec)

CR9052IEPE

200

None

Constant 2 mA

3.0 Hz to 20 kHz (

τ

= 0.5 Sec)

CR9052DC

7

V10

Constant 10 V DC

N/A

CR9052DC

5

None

Constant 5 V DC

N/A

CR9052DC

2

None

Constant 10 mA

N/A

CR9052DC

1

None

None N/A

Mult, Offset

Constant,Variable,
Array, Expression

A multiplier and offset by which to scale the measurement. A multiplier of one and an offset of 0
will return the measurement in millivolts.