Section 7. Measurement Instructions

7-47

The following program uses the SubScan to combine 2.5 kHz filtermodule measurements with 10 Hz
measurements on a CR9050 card.

' CR9052 example program #2
'
' Measure 2 channels on the CR9050 at 10 Hz on the +/- 5000 mV range.
' Measure six channels on the CR9052DC at 2.5 kHz on +/- 5000 mV range with 5-Volt excitation.
' Sample ratio is 2.5: top of pass band is 2.5 kHz / 2.5 = 1 kHz.
' CR9052 is in slot 8.
' Turn on flag 1 on to save instantaneous data to output table.


const stats_interval = 2

' time period over which to compute stats, in seconds


Public Flt_in (6)
units Flt_in = mV
Public Alg_in (2)
units Alg_in = mV
Public flag (1)

'Filter Module Filter Option

const SmplRat_2_5 = 2

'Fpass = Fsr/2.5 = 1/(T_scan*2.5)

'------------------- Data Tables -------------

DataTable (FiltData, flag (1), -1)

' save to final storage if flag (1) = True

Sample (6, Flt_in(1),IEEE4)
EndTable

DataTable (AlgData, flag (1), -1)

' save to final storage if flag (1) = True

Sample (2, Alg_in(1),IEEE4)
EndTable

'------------------- Program------------------

BeginProg
Scan (100, msec, 0, 0)

VoltDiff (Alg_in(1), 2, mV5000, 6, 1, False, 0, 0, 1.0, 0.0)

CallTable AlgData

SubScan (400, usec, 250)

'

VoltFilt (Destination, Reps, Range, FSlot, Chan, FiltOption, Excitation, Mult, Offset)

VOLTFILT

(Flt_in(1), 6, mV5000, 8, 1, SmplRat_2_5, 5, 1.0, 0.0)

CallTable FiltData

Next SubScan

Next Scan

SlowSequence

Scan (1, Sec, 0, 0)

Calibrate

' run calibrations for cr9050 measurements

BiasComp

Next Scan

EndProg