Cirrus Logic CS5373A User Manual
Page 35
![background image](/manuals/466398/35/background.png)
CS5373A
DS703F2
35
•
Full-scale Accuracy - Variation in the measured output voltage from the theoretical full-scale output voltage at
1x attenuation. The following formula is used to calculate full-scale accuracy:
•
Relative Accuracy - Variation in the measured output voltage from the theoretical attenuated output voltage at
each of the attenuation ranges. The following formula is used to calculate relative accuracy:
•
Full Scale Drift - The variation of the measured full-scale voltage across the specified temperature range.
•
Common Mode Drift - The variation in the measured common mode voltage across the specified temperature
range.
full scale accuracy =
measured full scale voltage - theoretical full scale voltage
theoretical full scale voltage
(
(
•100%
|
|
relative accuracy =
measured attenuated voltage - theoretical attenuated voltage
theoretical attenuated voltage (relative to the measured full scale voltage)
(
(
•100%
|
|