SERVICE

GyroMixer End User Guide

25

6.

Using mineral spirits or similar non-flammable liquid, remove the old

adhesive residue left on the table.

7.

Remove the adhesive paper on the back of the new pad.

8.

Carefully aligning the new pad, attach the pad to the upper table. Apply

pressure from the center outward and over the entire surface of the pad

until it is secure.

9.

Close the sliding door.

10.

Restore power to the mixer.

REPLACING THE SLIDING DOOR

1.

Unplug the mixer.

2.

Remove the rear access panel.

3.

Remove the top cover.

Figure 2. Door Covers

4.

Remove the two (2) screws and door stop from the right track. (Figure 3,

as viewed from back.)

TOP COVER

TRACK