,QFRPSOHWH FRPEXVWLRQ

Check the O

2

supply to the unit (30 bar). Use a combustion aid if necessary (see

also the applicable standards under ”Taking samples/sample preparation”.

&RYHU FDQQRW EH UHPRYHG IURP WKH GHFRPSRVLWLRQ YHVVHO

The O

2

filling / ventilation procedure is still running (see the status window in the

display).

:KHQ VXVSHQGLQJ WKH GHFRPSRVLWLRQ YHVVHO LQWR WKH PHDVXUHPHQW FHOO WKH

GLVSOD\ %RPE

p

LV QRW FKDQJHG WR 67$57

Check the following items:

“Stable” state not yet achieved (see status window).

The decomposition vessel was not removed when an experiment was inter-

rupted.

No experiment is prepared.

The maintenance menu is open.

Problem with the contact spring.

Ignition wire is defective.