-WERKE C 5000 control/duo-control Ver. 10 04.07

3DJH 

If it should become necessary to replace the O

2

filling piston (item 2), proceed as

follows:

c

Loosen the screws (item 6) with a blade-screwdriver.

d

Remove the centering ring (item 7) together with the filling head (item 5), the pres-

sure spring (item 3) and the piston (item 2)

Caution:

The seals (item 4 and item 8) are free

e

Push the pressure spring (item 3) onto the new piston and insert both parts into the

filling head. The remainder of the assembly takes place in the opposite order.

Caution:

During reassembly, make certain the filling head is in the

correct position (item 5). The seals (item 4 and item 8) should

be aligned on the opposite side of the inner cover.

$IWHU UHSODFLQJ WKH SLVWRQ DQG WKH VHDO GLVNV WKH GHFRPSRVLWLRQ YHVVHO PXVW

EH UHILOOHG ZLWK R[\JHQ XVLQJ WKH PDLQWHQDQFH PHQX PHQX LWHPV 2 )LOO  'H

SUHVVXULVH DQG WKHQ GHSUHVVXUL]HG LQ RUGHU WR WHVW WKH FRPSOHWH XQLW IRU

SURSHU VHDO

 5HSODFLQJ WKH 2



VHDO

If a leak is detected while filling the decomposition vessel with oxygen, the O

2

seal

on the filling piston must be replaced:

Remove the decomposition vessel from the measurement cell.

Activate the

2 VHDO

menu item from the

0DLQWHQDQFH

menu to extend the pis-

ton.

Remove the small orange seal from the extended piston (see Section 13.3,

item 11).

Insert the new seal (included with delivery) onto the piston.

Activate the

2 VHDO

menu item from the

0DLQWHQDQFH

menu again to retract the

piston again.

 'HFRPSRVLWLRQ YHVVHOV

For maintenance of the decomposition vessel, please read the C 5010/C 5012 Op-

erating Instructions.