On cleaning the decomposition vessel, see Section 10.4

To prepare the coded decomposition vessel, follow the steps listed below:

c

Screw off the cap screw and take off the cover with the aid of the handle.

d

Secure a cotton thread with a loop in it on the middle of the ignition wire.

e

Weigh in the calibration substance (about 1 g, 2 tables of certified benzoic acid; see

accessories), accurate to within 0.1 mg, and place in the crucible.