i

Place the crucible in the crucible holder.

j

Attach cotton thread

The decomposition vessel is fitted with a permanent ignition wire. In order to initiate

the combustion process, a cotton thread must be attached to the ignition wire. At-

tach the cotton thread as shown in the illustration.

Align the cotton thread with a pair of tweezers so that it is suspended into the cruci-

ble and touches the sample. This will ensure that the burning thread will ignite the

sample during the ignition process.

k

Close the decomposition vessel (C 5010 or C 5012).

Place the cover onto the lower section and push down until it presses against the

stop piece in the lower section.

C 5010

C 5012

Place the union nut onto the lower section and tighten by hand.

C 5010

C 5012

The decomposition vessel can now be filled.