IKA C 5000 control Package 2/12 User Manual
Page 75
IKA
-WERKE C 5000 control/duo-control Ver. 10 04.07
3DJH
e
Activate
6WDUW
. The measurement cell cover closes.
Each time 1000 ignitions have been performed using a given decomposition vessel,
the following message will appear:
1000 ignitions performed with Bomb x
Inspect decomposition vessel
or contact IKA service
This indicates that the decomposition vessel has reached a maintenance point and
that a safety check must be carried out. Confirm this message by pressing
7$%
and
2.
in sequence.
This message does not release the user from the responsibility of also continuously
checking the decomposition vessel for wear and carrying out safety inspections as
required.
The decomposition vessel is then filled with oxygen. Next, the inner vessel is filled
with water. As soon as the system begins with the experiment, the display shows a
graph of the change over time in the temperature of the inner vessel.
For a system with two measurement cells (duo-control) steps 1 – 2 are now possible
with the second measurement cell. In other words, the second measurement can be
started while the measurement in the first measurement cell is still running.
A measurement can also be manually ignited and stopped in order to carry out a
decomposition process quickly and safely. This requires use of an adiabatic or
isoperibol working method.
Ignite the sample using Menu
Î
Messungen
Î
Zündung (Menu
Î
Measurements
Î
Ignition). Ignition will occur within the next minute.
End the measurement using Menu
Î
Messungen
Î
Ende (Menu
Î
Measure-
ments
Î
Stop). The process will stop within the next minute.
f
The sample is ignited and the change in temperature of the inner vessel over time is
recorded.
After the end of the experiment, the system displays the results of the experiment