IKA C 5000 control Package 2/12 User Manual

Page 63

Advertising
background image

IKA

-WERKE C 5000 control/duo-control Ver. 10 04.07

3DJH 

1RWH

In all automatic calculations, 70 J is taken into account for

the electrical ignition energy.

6DPSOH QDPH

The software automatically assigns a sample number for each

measurement of the format

\PPGGQQ

, where

\

is the year,

PP

is

the month,

GG

the day and

QQ

a running number. It is easy with

sample numbers formed in this manner to select and work with

specific groups of measurements from the library.

If you select the option

8VHU GHI QDPH

under

0HQX &RQILJXUDWLRQ

6HWWLQJV

, you can assign your own numbers or names for meas-

urements (automatic numbering continues to run in the back-

ground, but is no longer taken into consideration).

If you have selected the option

/DVW

in addition under

0HQX &RQ

ILJXUDWLRQ 6HWWLQJV ([SHULPHQW LQLW

the number of the last ex-

periment appears as a suggestion for the current measurement. If

you do not edit this suggestion, the sample number will be the

same for all measurements!

Example of a

6DPSOH QDPH

= 6052401

1

Number of the year, 6 = 1996

05 Month, 0 … 12, 05 = May

24 Day of the month, here May 24

th

01 The running experiment number

6DPSOH

Any additional information on the sample. You can select letters

SURSHUWLHV

and characters from the character table with the

DUURZ NH\V

With

the

GHFLPDO SRLQW

key, the system accepts the selected characters

into the entry field (max. 40 characters).

8VHU

The name of the user (up to 8 characters). Entry as for

6DPSOH SURSHUWLHV

.

> @ &DOLEUDWLRQ

Mark this box with space bar for the system to use the experiment

for calibration.

2.

causes the system to accept entries in the dialog box.

g

The message

%RPE

p

appears from then on at the bottom of the screen. This

means that from now on, the decomposition vessel can be suspended in the meas-

urement cell cover.

h

If additional amounts of distilled water, solutions, etc. will be used in subsequent

combustion experiments, the same quantity of the same substance should be in-

cluded now in the decomposition vessel. The system should be calibrated in the

same state in which you intend to be working later. If the operating mode changes

(with/without an amount of water) the calibration should also be repeated.

7R LQFUHDVH WKH OLIH RI SDUWV VXEMHFW WR ZHDU DQG WHDU 2 ULQJV VHDOV HWF  ZH

UHFRPPHQG LQ JHQHUDO WKDW \RX DOZD\V ZRUN ZLWK D PHDVXUHG DPRXQW RI ZDWHU

LQ WKH V\VWHP

Advertising
This manual is related to the following products: