IKA C 5000 control Package 2/12 User Manual

Page 36

Advertising
background image

 6HWWLQJ XS DQG SODFLQJ LQ VHUYLFH

IKA

-WERKE C 5000 control/duo-control Ver. 10 04.07

3DJH 

f

Make sure that the oxygen supply line is connected to the C 5000 calorimeter as

described under Point

d

. Then connect the O

2

line to the laboratory oxygen supply

end.

Please refer to chapter 8.4 for details how to connect to oxygen supplies.

The oxygen supply line supplied is suitable for a pressure of max. 40 bar. It is 2 m

long, and the smallest permissible bending radius is 80 mm. The pressure-reducing

valve C 29 (accessory) is suitable for connecting the O

2

line to the oxygen supply. It

has a 1/4" BSP thread for connecting the hose. For use with an American pressure-

reducer with a 1/4" NPT thread, a suitable adapter for the oxygen line is included.

g

Place the C 5001 cooling system next to the measurement cell. Push the cooling

system all the way up to the measurement cell. The joining piece on the C 5001

cooling system fits into the opening on the measurement cell that lines up with it.

Fasten the two components in place by screwing in the two counter-sunk socket

bolts. Insert the hoses into the water connections of the measurement cell.