The system must be protected from mechanical bumps, vibrations, accumulations of

dust and corrosive ambient air during transportation and storage. It is also important

to observe that the relative humidity not exceed 80%.

In case of repair the device has to be cleaned and free from any materials which

may constitute a health hazard.

If you require servicing, return the appliance in its original packaging. Storage pack-

aging is not sufficient. Please also use suitable transport packaging.

 6HWXS ORFDWLRQ

To ensure high precision in measurements, a constant ambient temperature is re-

quired for the system. The following conditions must therefore be observed at the

setup location:

No exposure to direct sunlight.

No drafts (for example next to windows, doors, air conditioners).

A sufficient distance from heater blocks and other sources of heat.

Adequate circulation of air must be ensured to divert the system’s own heat.

The minimum distance between the wall and the rear side of the unit must not

be less than 25 cm.

The system must not have laboratory material such as shelves, cable sleeves,

ring leads, etc, built over it.

The room temperature must fall within the range of 20 – 25 °C.

The system must be set up on a horizontal surface.

To operate the system, the setup location must provide a power supply that con-

forms to the specifications on the rating plates of the system components, as well as

a supply of oxygen (99.95% pure oxygen, quality 3.5, pressure 30 bar) with the ap-

propriate pressure indicator. A shut-off valve for the oxygen supply must be in-

stalled. Observe the instructions on handling oxygen given in Section 1 "For your

safety”.