Install vapor tube and ptfe-seal – Heidolph LABOROTA 20 eco User Manual

Page 46

Advertising
background image

46

E

2. Install vapor tube and PTFE-seal

Unscrew large-size (1) and small (2) coupling rings. Lock driveshaft by turning drive lock (3)
CW to ease loosening and tightening of large-size ring (1).

Remove PTFE-adapter (4) with built-in PTFE-O-ring (5) and PTFE-seal (6)

Install backing ring (7) and vapor tube (8) in drive head. Put sandwich gasket (9) on

vapor tube coupling flange

Slide small ring (2) over coupling flange of distributor (10). Expand small split ring (11)

and slide over distributor (10) coupling flange, shoulder pointing forward

Secure assembly (10, 2, 11) tight with small coupling ring (2).

Slide PTFE seal (6) on vapor tube (8), sealing lip pointing inward

Never attempt to work with damaged or worn-out PTFE gasket (6). Not reaching
vacuum pressure wanted may damage your rotary evaporator. Inspect PTFE
gasket (6) in regular intervals and replace on condition.

Slide PTFE-adapter (4) with built-in PTFE-O-ring (5) into drive module.

Slide large-size coupling (1) over evaporator flask (12) coupling flange. Expand large split

ring (13) and slide over evaporator flask (12) coupling flange, shoulder pointing forward.
Set flask support tubes to 10 liter or 20 liter mark and lock with knobs (14); (I = 10 l, II =
20 l)

1

4 5

6

7

8

9

10

11

12

13

2

14

15

16

3