System re-start after an Stop-Mode situation requires system RESET. Make sure the
problem is fixed before resetting. See an expert, if required.

To Reset, turn Master switch OFF & ON.

28. Heating bath lift


Heating bath lift RAISES, as long as the Lift UP

key is held depressed.

Heating bath lift LOWERS, as long as the Lift DOWN

key is held depressed.

At upper and lower limits, lift motion stops automatically.

29.

Speed setting

The rotation is set using the "Speed" knob. On the display "Speed rpm" the speed is shown
in rpm (revolutions per minute).

By pressing the Start Stop button the rotation is started. The LED next to the Start Stop
button indicates that it is switched on. Press once again to stop the rotation.

CAUTION: for safety reasons, start rotation at low speeds. Rotational speed can be
increased during operation.
Never exceed a speed of 100 rpm, while working with a 20 l evaporation flask.

If rotation does not start with the START/STOP key, please check for illumination of one or
more of the Safety Cover, Flask Support or Rotation Stop LED’s.
If this occurs check if the enclosure is open, flask support not in one of its extreme positions,
or the drive drive lock engaged.

For more information refer to Troubleshooting Section.