Install condensers and expansion vessel – Heidolph LABOROTA 20 eco User Manual

Page 47

Advertising
background image

47

E

Locate evaporator flask on flask lift support tubes (15), tilt flask neck towards drive head,

center connecting flange in drive head, so that connecting flange rests against PTFE-O-
ring (5) located in PTFE-adapter (4).

Screw large-size ring (1) tight. When evacuating, double-check for firm seat

Return flask lift support tubes to initial position (16), CAUTION: basic position uses

electronic monitor circuit. Leaving this position will freeze all electric functions of the
LABOROTA 20 eco. This situation is reported by an LED on the control panel (flask
support).

3. Install condensers and expansion

vessel

Loosen thumbscrew (2), lift bracket (1)

completely and secure with
thumbscrew (2)

Place rubber side of washer (3) on

lower bracket (4), line up holes. Put
one more washer (3) with rubber side
pointing up on upper neck of condenser
or expansion vessel (5)

Plug lower neck of condenser or

expansion vessel (5) through cutout in
bracket (4) and washer (3), and tilt to
an upright position

Loosen thumbscrew (2) and slide

bracket (1) over upper neck of
condenser or expansion vessel. Make
vertical adjustment with bracket (4)

Rotate thumbscrew (2) tight when

condenser or expansion vessel are in
an exactly upright position.

3

1

5

2

4

3

6

3

Advertising