Error Messages and Warnings

No.

Displayed Message

Description and possible solutions

E04

TMP temperature control
defective (E025)

Short circuit in temperature sensor

E05

TMP temperature sensor
defective (E026)

Temperature sensor interrupted

E06

TMP frequency too high

Excess speed rate TMP

Replace TMP incl. TC 600.

Contact INFICON Service.

E07

TMP power supply unit
defective.

Power supply output voltage of the TC 600 too low.

24 V-powOperating Manualjinb80e1-her supply unit of Modul1000 - check output
voltage

Replace TMP incl. TC 600.

E08

TMP starting time error

After 15 minutes warm-up time, the speed of the TMP is lower than 1200 Hz.

Fore-vacuum pressure too high

Vacuum system is leaking

Bearing of turbo molecular pump defective

E09

TMP-connection TC600 - TMP
defective

Internal connection TC 600 to TMH 071 defective

Replace TMP incl. TC 600.

E10

TMP controller in TC 600
defective

Controller of TC 600 defective

Reset the controller with stopped pump (0Hz) using the power switch „ON/OFF“
of the leak detector.

Replace TMH incl.TC 600

E11

TMP wrong pump resistance
characteristics

Controller detects wrong pump resistance characteristics

Replace TMH incl.TC 600

Contact INFICON Service

E12

TMP motor control defective

Defect in motor output stage or motor control unit

Replace TMH incl.TC 600

Contact INFICON Service

W13

TMP unknown TMP-error

Error message from TC 600, for which there is error code specified in the software
of the leak detector.

The error code of the TC 600 converter is displayed.

W15

Leak rate is too high! Machine
was switched to Stand-by mode
to prevent HE-contamination!

The survey function „contamination“ is activated. A leak rate higher than the
adjusted value was detected.

Gross leak.

Switch off limit is set too low

Alarm delay time too short

W16

Service interval of leak detector
has expired!

Service interval of the leak test device has expired.

Carry out and acknowledge the service for the leak test device.

W17

Service interval of leak detector
has expired!

The service interval for the turbo molecular pump has expired! Period since the
service > 2 years

Replace the lubricant reservoir of the turbo molecular pump and acknowledge.