7.3

Maintenance schedule

Legend for the Maintenance
Schedule

•

IService segment ICustomer

•

IIService segment IICustomer with INFINICON training

•

IIIService segment IIIauthorized INFICON service technician

•

XPerform maintenance work according to operating hours or time

•

X

1

Perform maintenance work according to operating hours

•

X

2

Perform maintenance work according to time

•

1 depending on the environment and use

•

2 Relevant to the process

7.4

Maintenance intervals

The maintenance schedule for the Modul1000 can be subdivided in 4 maintenance
groups.

•

1500 hours maintenance

•

5000 hours maintenance

•

Maintenance every two years

Assembly

Maintenance tasks

Operating hours/Years

Service

segment

Spare

Standards

1500

5000

15000

Part No.

1/4

1

2

3

Vacuum system

Valve block

Clean the valves, replace seals for valves

X

1

III

200000594

Take apart the valve bloc and clean it

X

1

III

200002002

Replace the filter of the flood gas line

1

X

1

I, II,III

200000683

Adjust the Pirani

X

III

Recalibrate the internal test leak

X

2

TMH 071

Replace the lubricant reservoir

X

2

I, II and III 200000577

Change the bearings (recommended)

X

2

III

Electric

Fans assembly

Replace the filter inserts of the cover

1

X

1

I

200001552

Clean the fan, chassis and turbo-molecular
pump

1

X

1

I