9 replace vent filter, Replace vent filter – INFICON Helium Leak Detector Modul1000 User Manual
Page 86

7-14
Maintenance tasks
jinb
80
e1
-h
Op
er
at
in
g M
a
nua
l
(1
01
2)
7.9
Replace Vent Filter
The flood filter is checked and replaced in the course of the 5000-hours mainte-
nance. In case of extreme operating conditions shorter maintenance times could be
reasonable.
Required Tools
•
Allen key SW 8
Required material
•
Replacement filter (2 pcs.) Order No.: 200 000 683
Access to the flood filter is only possible when the cover of the Modul1000 has been
removed. Follow the instructions in
.
Notice:
Please observe the safety instructions in this chapter.
•
To dismantle the vent filter, disconnect the hose connections of the filter from the
quick-connecting couplings. By pressing on the outside ring on the quick
connecting couplings, the hose can be pulled off. Please find the mounting
positions in the following
Fig. 7-8
.
Fig. 7-8 Flood filter replacement in Modul1000
Item Description
Item Description
1
Vent filter
2
Hose routing
a
1
a