Operation

6.1

Switching ON

Install the device as described in the Installation chapter. Connect the power cable.

Switch the device ON. The mains switch and the connection for the power cable are
at the back of the device.

After the mains switch has been actuated the devices runs up automatically.

In run-up mode (

≤ 3 Min.) the following items appear on the display:

Speed of the turbo-molecular pump

Fore-vacuum pressure

State of emission

Active filament

A bar graph which shows the run-up progress

After the run-up has been completed, the Modul1000 is in „Standby“ mode.

6.2

Status LED

The LEDs show the operating mode of the device:

Operating mode:

LED, green

LED, yellow

Run-up

Flashes

Flashes

Standby/Vent

On

Off

Evacuate

On

Flashes slowly

Measuring

On

On

Calibrating

Flashes synchronously

Flashes synchronously

Error / Warnings /
Maintenance messages

Off

Flashes fast