6-21

Operation

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

→

Reference measurement

In this menu, the reference measurement can be started.

Main menu

→

Settings

→

Vacuum settings

→

Commander Functions

→

In the menu item „Commander Functions“, all parameters of the commander function
can be configured.

→

Commander Timing

The individual process times can be set here.
The adjustment range is from 0.1 to 95 seconds

t_A evacuation time

Maximum time to achieve the set sample pressure "p_B evacuation pressure"

Default setting: 30 seconds

t_B zero delay

Time delay until ZEROING (if zero "enabled") or time in which I·ZERO is carried out
(if I·ZERO has been activated);

Default setting: 10 seconds

t_C filling time

Maximum time to achieve the filling pressure "p_C Filling pressure" in the unit under
test

Default setting: 30 seconds

t_D depressurising time

Maximum time to achieve the set depressurising pressure "p_B evacuation pres-
sure"

; factory setting: 30 seconds

t_E flooding time

Maximum time to achieve the set flooding pressure p_A gross leak test

; factory setting: 10 seconds

t_F readiness for measuring

Maximum time to achieve the Measuring mode

; factory setting: 10 seconds

t_G measuring time

Time, after which the leak rate signal has been balanced.

; factory setting: 10 seconds