10-2

Error Messages and Warnings

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

W18

Service interval of fan filter has
expired!

The service interval for the fan filter chassis has expired!

Replace the fan filter and acknowledge the service.

W21

EEPROM time-out for write
instruction

Write instruction from the MC68 to the EEPROM was not acknowledged.

E-EPROM defective.

Defect in cross wiring

MC 68 is defective

W22

Overflow of EEPROM
parameter queue

Software problem.

Please contact the INFICON Service!

E23

External supply voltage (24 V)
too high

Voltage monitoring input AD24 A/B > 11.5 V.

External power supply on one of the 24 V outputs of the terminal connector of the
interface board.

E24

External supply voltage (24 V)
too low

Voltage monitoring input AD 24 A/B < 2.5 V.

Check the fuses F1 to F4 on the interface board. At least 2 fuses failed.

E25

Reduced valve voltage too low
(<7 V)

Reduced valve voltage on E/A board (<7V)

E/A board is defective.

MC 68 control board is defective..

E26

Monitoring voltage F3, F4

Voltage monitoring for fuses F3, F4 AD 24 A < 5,6 V.

Fuse F3 interface board (SSK) defective.

E27

Monitoring voltage F3, F4

Voltage monitoring for fuses F3, F4 AD 24 A < 7.8 V.

Fuse F4 interface board (SSK) defective.

W28

Real time clock has been reset!
Please enter date and time!

Real time clock has been reset!

Battery at MC68 is discharged or faulty.

MC68 has been replaced.

E29

24V supply for fans is defective.

Power supply unit for fans is defective < 20 V.

Fuse F11 on wiring backplane has blown.

Fan polarity has been interchanged.

W31

The offset voltage of the pre-
amplifier is too high. (>5 mV)

Offset voltage of pre-amplifier without emission > 5 mV.

The pre-amplifier is faulty.

Defective voltage supply of pre-amplifier.

Defective MC 68 control board.

W32

Pre-amplifier temperature is too
high. (>60°C)

The temperature of the ambience is too high.

Heat accumulation due to unfavourable placement of the module.

Air filter dirty and have to be changed.

The temperature sensor in the pre-amplifier is defective.

Defective MC 68 control board.

W33

Pre-amplifier temperatre is too
low. (<2°C)

Ambient temperature is too high.

The temperature sensor in the pre-amplifier is defective.

Defective MC 68 control board.