5.4.1

Design of a leak detector system

Design of a leak detector system with Modul1000- - -Commander function is mainly
the same as that of the conventional integrated leak test system.
The vacuum chamber is evacuated by the Modul1000, optionally also in partial flow
mode (

Fig. 5-1

Partial flow mode

). The test sample in the vacuum chamber is

connected to the helium filling station through the chamber walls and is filled with
helium after the chamber has been evacuated.

Due to the difference between the helium pressure in the test sample and the
vacuum in the vacuum chamber, the helium of a leaky part flows out of the leak into
the vacuum chamber and into the leak test device where it can be measured as gas
leak.

Fig. 5-2 Leak detector system

Item

Description

Item

Description

V30

Valve evacuation (test sample)

1

Vacuum chamber

V31

Flood valve (test sample)

2

Test sample

V32

Valve for helium recovery

3

Pressure gauge

V33

Filling valve (test sample)

4

Vacuum valve (test sample)

V34

Emergency valve (de-energized
open)

5

Fore-pump (Modul1000)