7 fuse replacement, 1 overview on electrical fuses, Fuse replacement – INFICON Helium Leak Detector Modul1000 User Manual

Page 82: Overview on electrical fuses

Advertising
background image

7-10

Maintenance tasks

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

7.7

Fuse replacement

7.7.1

Overview on Electrical Fuses

Notice:

Please note that this work is only performed by skilled electricians, since
generally a cover of an electric module may have to be opened.

Mains switch housing chassis:

Designation

Technical data Fuse for

Mains switch

2 x T 6.3 A

Power supply unit fuse (2-phase OFF)

Wiring plane:

Designation

Technical data Fuse for

F10

T 6,3 A

Power supply unit TC 600

F11

T 0,8 A

Fans

Power supply unit ZWS240PAF-24/TA:

Designation

Technical data Fuse for

F1

F 6.3 A

Fuse power supply unit

I/O PCB:

Designation

Technical data Fuse for

F1; F2

T 0,8 A

not used

F3

T 0,315 A

not used

Circuit board MSV:

Designation

Technical data Fuse for

F1

T 2 A

24 V main fuse for MSV card

F2

T 3,15 A

Anode heater (no use)

F3

T 1 A

15 V; +5V DC/DC transformer

F4

M 0.032 A

Anode- cathode voltage (85 V)

Interface board:

Designation

Technical data

F1

T 1 A

24 V CONTROL UNIT

F2

T 0,8 A

24 V REMOTE CONTROL; PC RS232

F3

T 0,8 A

24 V PRESSURE GAUGE, PLC IN

F4

T 1,6 A

24 V PLC OUT; VALVES; ACESSORIES

F5

T 1,0 A

VALVES V30...V33 supply voltage (max.
30 V)

F6

T 1,0 A

VALVES V34...V37 supply voltage (max.
30 V)

Advertising