6.8

Calibration in Auto Leak Test

There are four different ways of calibration:

Internal automatic

Internal manual

External automatic

External manual

For the external calibration, a separate test leak is required, which is mounted on the
chamber.

In Partial Flow mode, always perform an external calibration.

Sequence of calibration:

1

open the test leak or it opens automatically,

2

evacuate,

3

acknowledge steady-state signal when calibrate manually,

4

autotune (Calibration to maximum sensitivity),

5

vent to atmospheric pressure,

6

evacuate, after the measuring time is over, take the signal with open test leak,

7

vent to atmospheric pressure,:

8

close the test leak, evacuate, after the measuring time take the background signal,

9

vent to atmospheric pressure.

6.9

Calibration in Commander mode

During the calibration, the test gas supply is not enabled.
There are four different ways of calibration:

Internal automatic

Internal manual

External automatic

External manual

For the external calibration, a separate test leak is required, which is mounted on the
chamber.

The calibration procedure is the same as it is for the vacuum mode.