Sniffer tip

5.3.1

Replacing the sniffer tip

Fig. 5: Sniffer tip

1 Switch the device off.

2 Loosen the cap nut until the sniffer tip can be pulled out.

3 Insert the new sniffer tip into the sniffer probe in such a way that the guide pin runs into

the groove of the sniffer probe opening.

4 Tighten the cap nut.

5 Calibrate the device,

see "6.4.6.1 Time and type of calibration", page 43

.

Further sniffer tips

see "10.1 Accessories and spare parts", page 67

.

To replace the filter

see "8.2 Sniffer line", page 63

.

Material damage from pollution

Particles in the air intake can destroy the sniffer line.

Always switch off the device before every installation!

Make sure that no particles enter into the air intake when replacing sniffer probe
parts.

a

Filter block

b

Guide pin

c

Cap nut

d

Sniffer tip

e

Filter holder