3 display on the sniffer line, Display on the sniffer line, See table 7 on – INFICON HLD6000 Refrigerant Leak Detector User Manual
Page 33

6 Operation
33
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
6.2.3
Display on the sniffer line
An LED is mounted on the sniffer handle where you can read the different operating
modes,
.
Fig. 12: Sniffer line display
a
see "6.4.6.1 Time and type of calibration", page 43
b
Status LED
c
Filter holder on the sniffer tip
d
Designation of the gas that the sniffer line sniffs.
Table 7: Signals of the LED on the handle grip
Operating status
LED
Not connected
Off
No communication
Blue, flashing
Run up
Blue
Standby
Blue, flashing
Measuring mode
Green
Leakage rate >40 % of the setpoint
Yellow
Leakage rate >100 % of the setpoint
Yellow, flashing
Calibration
Blue, flashing
Error/Warning
Red, flashing
Error/Warning and leakage rate > 40 % of the setpoint
Red/green, flashing
Error/Warning and leakage rate > 100 % of the setpoint
Red/yellow, flashing