Display on the sniffer line

An LED is mounted on the sniffer handle where you can read the different operating
modes,

see Table 7 on Page 33

.

Fig. 12: Sniffer line display

a

Calibration key,

see "6.4.6.1 Time and type of calibration", page 43

b

Status LED

c

Filter holder on the sniffer tip

d

Designation of the gas that the sniffer line sniffs.

Table 7: Signals of the LED on the handle grip

Operating status

LED

Not connected

Off

No communication

Blue, flashing

Run up

Blue

Standby

Blue, flashing

Measuring mode

Green

Leakage rate >40 % of the setpoint

Yellow

Leakage rate >100 % of the setpoint

Yellow, flashing

Calibration

Blue, flashing

Error/Warning

Red, flashing

Error/Warning and leakage rate > 40 % of the setpoint

Red/green, flashing

Error/Warning and leakage rate > 100 % of the setpoint

Red/yellow, flashing