Evaluating measured data

A file with measured date is constructed as follows:

For example

// Record file: \L0000001.txt

// Created by HLD6000CU V0.11.02.18681

// HLD6000CU Ser.-No.: 00000000000

// HLD6000 Ser.-No.: 00000000000

// HLD6000MB V0.22.06(1.04.00)

// Probe V1.00

// Probe Ser.-No.: HLD5000 probe

// Probe Type: SMART (R134A)

// IO1000 V0.05.00(0.02.02)

// IO1000 Ser.-No.: 00000000000

// BM1000 not connected

// Start time: 23.06.2014 08:58:25

Time Leakrate[g/a] Status

0.0 1.82E-02 MEASURE

0.5 1.82E-02 MEASURE

1.0 4.16E-03 MEASURE

1.5 1.29E-02 MEASURE

2.0 0.00E+00 MEASURE

2.5 1.02E-02 MEASURE

3.0 1.75E-03 MEASURE

3.5 3.43E-03 MEASURE

4.0 3.43E-03 MEASURE

4.5 0.00E+00 MEASURE

5.0 8.20E-03 MEASURE

5.5 1.71E-02 MEASURE

6.0 1.52E-02 MEASURE

In the example shown the data record started on 06/23/2014 at 8:58:25. From this start
time onwards a measurement result was recorded every 500 milliseconds.

1.82E-02 means 1.82 x 10

-2

as the leakage rate per year.

The following status entries are possible:

6.4.8.3

Transferring measured data from the internal memory to a USB stick

It is possible to transfer measured data on the internal memory to a connected USB stick.

1 Connect a USB stick with the USB port on the device.

2 Select "

> Recorder > Copy".

3 Select the files you wish to copy.

4 Confirm your selection using

.

Status entry

Meaning

UNKNOWN

Unknown, as, e.g., there is no communication

RUNUP

Run up

STANDBY

Standby

MEASURE

Measuring

ERROR (xxx)

Error (error number)

WARNING (xxx)

Warning (warning number)

CAL

Calibration