7 Warning and error messages

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

HLD

6000

, k

ina4

3

e1-

a

, 140

8

E133

EEPROM error in sniffer line

The EEPROM in the sniffer line is defective

Contact INFICON customer service.

E134

Protocol error when communicating
with the sniffer line

The interface to the sniffer line does not function
correctly.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

E135

Checksum error when communicating
with the sniffer line

– The interface to the sniffer line does not function

reliably.

– Possible electrical sources of disturbance close

by.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– Eliminate sources of disturbance

– If the problem still exists, contact INFICON customer

service.

E136

No response from the sniffer line

The interface to the sniffer line does not function
correctly.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

E137

Resetting the sniffer line

– The sniffer line has reset itself.

– Possible electrical sources of disturbance close

by.

– The sniffer line is not connected correctly.

– Eliminate sources of disturbance

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

E138

Checksum error in EEPROM on the sniffer
line

The EEPROM in the sniffer line has false data or is de-
fective.

Contact INFICON customer service.

E139

Empty EEPROM in the sniffer line

The EEPROM in the sniffer line has no data or is de-
fective.

Contact INFICON customer service.

W140 Acceleration of the probe was perma-

nently too high within the last 5 minutes

– The sniffer line is not lying still.

– Defect in the sniffer line.

– Place the sniffer line on a steadier surface.

– Contact INFICON customer service.

W151 No connection to operating unit

Internal connection problems between the basic
unit and the operating unit.

Contact INFICON customer service.

W163 COOL-Check is not connected

COOL-Check is not connected or not connected cor-
rectly.

– Connect a COOL-Check, otherwise confirm the warn-

ing and calibrate externally.

– Check the COOL-Check connection.

W164 Checksum error in COOL-Check

– The COOL-Check is not connected correctly.

– The COOL-Check or the basic unit is defective.

– Check the COOL-Check connection.

– Check the connection of the COOL-Check to basic unit

(separate and reconnect; try another COOL-Check, if
possible).

– If the problem still exists, contact INFICON customer

service.

Table 10: Warning and error messages (Contin.)

No. Message

Possible

error

sources

Troubleshooting