7 Warning and error messages

59

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

HLD

6000

, k

ina4

3

e1-

a

, 140

8

6xx Calibration errors

W630 Calibration request

Calibration is old or no longer applicable.

– Recalibrate the device.

– Select a suitable interval in the menu for the calibra-

tion request.

W631 Light barrier is overmodulated

The light barrier in the calibration opening receives
too much light.

Ensure that no light or direct sunlight shines into the
calibration opening.

W632 Light barrier blocked during run-up

Dust has accumulated in the calibration opening
and is interrupting the light barrier.

– Switch the device off.

– Blow out the calibration opening with clean com-

pressed air.

– Restart the device.

If this does not work, then calibrate externally with the
COOL-Check or an external calibration leak.

7xx Temperature errors

W710 Motherboard temperature is too high

– The ambient temperature is too high.

– The fan is defective or blocked.

– Decrease the temperature around the device.

– Clean the ventilation openings or replace the filter

plates.

– Contact INFICON customer service.

E711

Motherboard temperature is far too high – The ambient temperature is too high.

– The fan is defective or blocked.

– Switch off the device and allow it to cool down.

– Decrease the temperature around the device.

– Clean the ventilation openings or replace the filter

plates.

– Contact INFICON customer service.

W730 COOL-Check temperature is outside the

range

– The basic unit is standing on a hot surface.

– The temperature around the main unit is too

high or too low.

– Remove the device from the hot surface.

– Decrease or increase the temperature around the de-

vice.

9xx Maintenance information

W902 COOL-Check is almost empty

– The COOL-Check is empty.

– The basic unit is set to a wrong date.

– Change the COOL-Check.

– Set the basic unit to the current date.

W903 COOL-Check is empty

– The COOL-Check is empty.

– The basic unit is set to a wrong date.

– Change the COOL-Check.

– Set the basic unit to the current date.

W904 Replace the filter holder on the sniffer tip The filters on the sniffer tip should be replaced.

– Replace the filters.

– Select a suitable interval in the "Filter change re-

quest" menu.

Table 10: Warning and error messages (Contin.)

No. Message

Possible

error

sources

Troubleshooting