1.4 Installation

W

ARNING

: Care should be exercised to route SQC-222 cables as far as practical from

other cables that carry high voltages or generate noise. This includes other line voltage
cables, wires to heaters that are SCR-controlled, and cables to source power supplies
that may conduct high transient currents during arc down conditions.

Rack
Installation

The SQC-222 occupies a 5.25” high, half-rack space. An
optional installation kit is available to adapt to a full rack (see
Chapter 4). Install the unit in a 19” rack with the supplied
hardware.

Power
Connection

The SQC-222 automatically detects mains voltages of 100-120
and 200-240VAC, 50/60Hz.

W

ARNING

: Verify that the power cable provided is connected to

a properly grounded mains receptacle.

Sensor Input
Connections

Connect the BNC cables and oscillators from your vacuum
chamber feedthrough to the desired SQC-222 sensor inputs.
See the previous section for cabling details.

Source Output
Connections

Connect the BNC cables from the SQC-222 output connectors
to your evaporation supply control input. Consult your Power
Supply operator’s manual for control input wiring instructions.

Digital I/O
Connections

Refer to Appendix C for details on wiring digital I/O to the SQC-
222 Relay I/O connectors. Appendix C also covers I/O wiring
with the optional PLC.

Computer
Connection

If you would like to use the Windows software to collect data or
program the SQC-222, attach a 9 pin straight-thru cable from
the RS-232 connector to your computer’s serial port. A cable
is supplied with the SQC-222.