Chapter 5 Communications

5-6

U0

A

Shutter is closed.

Command: Y

Parameters: None.

Description: Read the Power-Up Reset flag. The Power-Up Reset flag is set during

boot-up of the unit and stays set until read through the RS-232 interface.
After the flag is read, it is reset and will not be set again until the unit is
power cycled.

Example:

Y

A1

Power-Up Reset flag is set.

Y

A0

Power-Up Reset flag is reset.

Command: Z

Parameters: None.

Description: Set all Film and System parameters to defaults.

Note that this command can take over 1 second to complete

Example:

Z

A

All Film and System parameters are set to

defaults.