CrystalSix Operating Manual

Chapter 2

Sensor Installation

Successful operation of any crystal sensor depends on proper placement,
compatibility of its construction with its operating environment and connection to
proper utilities.

NOTE: The sensor head, water tubes, cable, etc. should be clean and grease free

when installed in the vacuum chamber. These parts should be handled
while wearing clean nylon gloves. If parts do become contaminated, clean
them thoroughly using a suitable solvent to avoid outgassing and
excessive peeling of deposition material from the sensor’s surfaces.

CAUTION

Incorrect rotation direction will cause permanent damage
rendering transducer inoperative. Refer to

Figure 1-2 on

page 1-5

for proper direction of rotation.

2.1 Crystals in the CrystalSix Sensor

2.1.1 Handle the Crystal with Care

Always use clean nylon lab gloves and clean plastic tweezers when handling the
crystal. Handle the crystals only by their edges. Anything that comes in contact with
the crystal surfaces may leave contamination, which may lead to poor film
adhesion. Poor film adhesion will result in high rate noise and premature crystal
failure.

CAUTION

Do not use metal tweezers to handle crystals. Metal
tweezers may chip the edge of the crystal.