3 assembly instructions – INFICON CrystalSix Sensor User Manual

Page 43

Advertising
background image

4 - 7

IP

N 07

4-

15

5L

CrystalSix Operating Manual

CAUTION

When removing the top plate make certain the connector
terminal does not become damaged.

10

Remove the compression spring (item #28).

11

Remove the carousel assembly (item #5)

CAUTION

When removing the carousel make certain the 0.126"
diameter ball (item #27) does not become lost.

4.4.3 Assembly Instructions

In the following steps, reference

Figure 1-3 on page 1-6

and

Figure 4-2 on page

4-6

.

1

Place the 0.126" diameter ball (item #27) in the countersunk hole of the sensor
body (item #11).

2

Attach the corrugated springs (item #31) if not already attached to the sensor
body (item #11), as illustrated in detail "F" of

Figure 1-3 on page 1-6

, utilizing

the 0-80 pan head screws (item #45).

CAUTION

Be extremely careful when handling the corrugated
springs to avoid distortion. The successful operation of
this apparatus is dependent on this component's
integrity.

3

With the free end of each corrugated spring outside the sensor body, install the
carousel assembly (item #5) so that it rests on the .126" diameter ball. Insert
the corrugated springs into the space between the carousel (item #5) and the
body (item #11). Make certain that the corrugated springs are not damaged
upon installation.

4

Rotate the carousel clockwise until the alignment hole in the carousel lines up
with the alignment hold of the body. Insert the 0.125" diameter alignment pin of
the alignment tool assembly (750-254) provided. Refer to

Figure 4-2 on page

4-6

.

Advertising