1 alignment instructions – INFICON CrystalSix Sensor User Manual

Page 40

Advertising
background image

4 - 4

IP

N 07

4-

15

5L

CrystalSix Operating Manual

CAUTION

Incorrect rotation direction with the stops disabled will
cause permanent damage rendering transducer
inoperative. Refer to

Figure 1-2 on page 1-5

for proper

direction of rotation.

The unit, when properly aligned, will index the crystal to a position that is centered
within the aperture provided in the heat shield. The unit must be aligned if this
condition does not exist.

Reference

Figure 1-3 on page 1-6

.

Equipment Required:

Š

5/64" allen wrench (Included in 750-268 Kit)

Š

1/16" allen wrench (Included in 750-268 Kit)

Š

3/32" allen wrench (Included in 750-268 Kit)

Š

Alignment tool assembly 750-254-G1 (Included in 750-268 kit)

Š

Regulated air supply 80-90 PSIG (5.5 bar - 6.2 bar) [550 kPa - 620 kPa]

NOTE: 75 PSIG (5.2 bar) [520 kPa] for alignment purposes only)

4.4.1 Alignment Instructions

With the required equipment at hand, proceed as follows:

1

Temporarily connect an air supply to the bellows assembly supply tube.
Regulate the air pressure to 75 PSIG (5.2 bar) [520 kPa].

2

Remove the actuator cover (item 15) and related hardware.

3

Insert the 0.125" diameter alignment pin (of the 750-254-G1 assembly) in the
alignment hole (if not already installed) as shown in

Figure 4-2 on page 4-6

.

The carousel may need to be rotated to achieve the alignment condition shown
in

Figure 4-2

.

4

Loosen the set screws (two set screws per item) of the ratchet (item #22) and
the stop ratchet (item #38).

5

Loosen the #2 fastener (item #42) that attaches the detent (item #41) to the
bellows assembly item #19). Do NOT remove detent.

6

Actuate the bellows assembly by applying 24 volts to the leads of the solenoid
valve of the solenoid valve assembly. Consult

Figure 3-1 on page 3-4

for wiring

assignments. The air pressure must be sustained.

Advertising