2.6 CrystalSix Sensor Installation when used

with an IC/4 or IC/4 PLUS

2.6.1 Programming the Relay Outputs

1

From the program menu select Source/Sensor (F4).

2

Select Sensor board 1 or Sensor board 2.

3

Enter the edit mode. (F5)

4

Select Sensor type six (6) for the CrystalSix Sensor. This will enable the
CrystalSwitch output.

5

Select a CrystalSwitch output (Relay 1-8 or 1-16) depending upon which I/O
board is used.

6

Turn the instrument’s power “OFF”, then “ON”, to ensure that the change takes
place permanently.

2.6.2 Wiring the Relay Outputs with Relay I/O Boards

755-122-G1 or 755-122-G1/G2

1

On the relay connector, see

Figure 2-7 on page 2-14

, connect the leads of the

solenoid valve across the one side of a 24 volt supply (see listing) and one side
of a relay (1-16 whichever was programmed). Connect a jumper between the
second side of the 24 volt supply and the second side of the selected relay.
Refer to

Figure 2-5 on page 2-9

for a typical installation utilizing relay 1 to

perform the crystal advance function for the relay I/O board.

Table 2-1 Wiring Connectors

(P4)

Top Relay Connector

(P5)

Input Lines

Relay

Pins

Input

Pins

1

24, 28

1

1, 20

2

23, 27

2

2, 21

3

22, 26

3

3, 22

4

13, 19

4

4, 23

5

12, 18

5

5, 24

6

11, 17

6

6, 25

7

10, 16

7

7, 26

8

3, 7

8

8, 27