2 disassembly instructions, Own in, Figure 4-2 – INFICON CrystalSix Sensor User Manual

Page 42

Advertising
background image

4 - 6

IP

N 07

4-

15

5L

CrystalSix Operating Manual

Figure 4-2 Alignment Pin Installation

4.4.2 Disassembly Instructions

In the following steps, reference

Figure 1-3 on page 1-6

and

Figure 4-2 on page

4-6

.

1

Remove the two #4 fasteners that secure the actuator cover (item #15) to the
top plate (item #10).

2

Remove the actuator cover. Remove the heat shield (item #8) by gently pulling
on the outside of the shield, the shield should snap off.

3

Remove the extension spring from the pawl and actuator stem assembly (item
#24).

4

Remove the two #4 fasteners (item #20) that secure the bellows assembly
(item #19) to the top plate (item #10).

5

Carefully remove the bellows assembly. The pawl and actuator stem assembly
(item #24) and the detent (item #41) will be removed with the bellows assembly.

6

Using the 1/16" allen wrench, loosen the set screws (item #33) that secures the
ratchet (item #22) and the stop ratchet (item #38) to the carousel assembly
(item #5). Remove the ratchet (item #22) and the stop ratchet (item #38).

7

Remove the three #4 fasteners (item #14) that secure the top plate (item #10)
to the sensor body (item #11).

8

Remove the top plate (item #10).

9

Remove the Teflon washer (item #39).