Attach the detent and related hardware (items 40, 42-44) to the new bellows
assembly. Do not tighten. Position the detent (item #41) so that it engages the
ratchet (item #22).

Unit is now ready to be aligned. Refer to

section 4.4.1 on page 4-4

for alignment

instructions.

Figure 4-3 Setting the Stop Ratchet

Figure 4-4 Setting the Detent