Cygnus Editor Operating Manual

files may have different limits they are left blank if more than one file is selected. If
the ranges selected exceed the actual limits of one of the files in a group, the points
that fall within the specified ranges are reported and no error is displayed.

Click on the Data Format tab to display the Data Format page, shown in

Figure

4-19

, used to select the formatting options and the type of data for the report.

NOTE: It is important to note here that even though the data file contains four data

parameters (Rate, Rate Deviation, Power, Thickness) for each channel, a
Report can only contain either one data parameter for each Channel or all
data parameters for all channels. The Data Unit choice is used to choose
which parameter is being reported.

Figure 4-19 Generating a Report - Setting the Data Format

The data can be oriented in Columns or Rows. Selecting Columns places all the
measurements for each parameter in a column with one scan per row. Selecting
Rows generates a series of parameter-measurement pairs, grouped by scan. The
delimiter between columns can be selected as either Tabs or Commas.