Process menu – INFICON Maxtek PM-700 Plating Monitor User Manual

Page 40

Advertising
background image

PM-700 SERIES PLATING MONITOR

32

4.5.3.2

Process Menu

Numonic Name/Description

Minimum Maximum

Default

Process Process

The plating monitor is
capable of holding up to
four different processes.
This parameter allows
selection of one of the four
processes.

Number 1

Number 4

Number 1

EndPoint
Setting

Endpoint Setting

Allows setting of the plating
endpoint thickness. This is
the desired thickness to be
plated on the substrate.
When the monitor displayed
thickness reaches this
thickness setting, the
monitor displayed thickness
will begin flashing rapidly,
informing the user that the
plating process is done.
Also, the Thickness
Achieved Output will goes
high (true). This output
could be used to turn on
other remote of alarming
device.

0.000 999.9 10.00

Min Rate
Alarm

Minimum Rate Alarm

Allows the user to set the
minimum allowable plating
rate. When the plating rate
drops below this set value,
the displayed rate will flash
rapidly. Also there is a Low
Rate Alarm Output that will
come on at this time. It
could be used to turn on
other remote alarming
device.

0.000 999.9 0.000

Max Rate
Alarm

Maximum Rate Alarm

Similar to the Minimum
Rate Alarm parameter, this

0.000 999.9 9.999