PM-700 SERIES PLATING MONITOR
32
4.5.3.2
Process Menu
Numonic Name/Description
Minimum Maximum
Default
Process Process
The plating monitor is
capable of holding up to
four different processes.
This parameter allows
selection of one of the four
processes.
Number 1
Number 4
Number 1
EndPoint
Setting
Endpoint Setting
Allows setting of the plating
endpoint thickness. This is
the desired thickness to be
plated on the substrate.
When the monitor displayed
thickness reaches this
thickness setting, the
monitor displayed thickness
will begin flashing rapidly,
informing the user that the
plating process is done.
Also, the Thickness
Achieved Output will goes
high (true). This output
could be used to turn on
other remote of alarming
device.
0.000 999.9 10.00
Min Rate
Alarm
Minimum Rate Alarm
Allows the user to set the
minimum allowable plating
rate. When the plating rate
drops below this set value,
the displayed rate will flash
rapidly. Also there is a Low
Rate Alarm Output that will
come on at this time. It
could be used to turn on
other remote alarming
device.
0.000 999.9 0.000
Max Rate
Alarm
Maximum Rate Alarm
Similar to the Minimum
Rate Alarm parameter, this
0.000 999.9 9.999