4 handling microtome knives/blades, 5 knife guard/handwheel lock, 6 motorized sectioning – Leica Biosystems CM3050 S User Manual

Page 13: 7 defrosting/handling frozen tissue, Safety

Advertising
background image

13

Leica CM3050 S – Cryostat

2.3.4 Handling microtome knives/blades

• Danger!

--> Microtome knives and disposable blades

have extremely sharp cutting edges and

can cause serious injuries. Therefore:

• Handle knives / blades with utmost care.

• Never leave any knives / blades in unpro-

tected places!

• Never place a knife, no matter where,

with the cutting edge facing upwards.

• Never try to catch a falling knife.

• Always insert the specimen before in-

serting the knife.

2.3.5 Knife guard/handwheel lock

• Always cover the cutting edge with the knife

guard and lock the handwheel:

• Prior to doing any work on knife and / or

specimen.

• Prior to exchanging specimens.

• During work breaks.

2.3.6 Motorized sectioning

• Do not interrupt sectioning / trimming by set-

ting the sliding potentiometer to zero speed!

--> This does not really switch the sectioning

function off - it only operates at ‘0’ speed.

If the sliding potentiometer is acciden-

tally moved, the instrument will resume

sectioning immediately (risk of injury)!

• During motorized sectioning, always center

the handwheel grip!

2.3.7 Defrosting/Handling frozen tissue

• Never leave specimens unattended in the

cryochamber over an extended period of

time!

--> In case of power failure or instrument

failure, or during the automatic defrost cy-

cle, tissue material can be destroyed.

• During the defrost cycle the cryochamber is

partially warmed. - Therefore:

• Remove sensitive specimens from the

chamber prior to defrosting.

• If automatic defrosting is programmed

to take place during the night, remember

to remove all specimens from the cryo-

chamber prior to leaving work.

2.3.8 Frozen parts of the instrument and

frozen accessories

• Prolonged contact of bare skin to frozen sur-

faces of the instrument or to frozen acces-

sories (specimen discs, knife holder, shelves

etc.) can cause frostbite.

• Wear protective gloves.

2. Safety

Advertising
This manual is related to the following products: