2 potential problems - causes and remedies, Potential problems - causes and remedies, Troubleshooting, applications tips – Leica Biosystems CM3050 S User Manual

Page 48

Advertising
background image

48

Instructions for Use V 1.7 RevC - 08/2014

6.

Troubleshooting, applications tips

6.2 Potential problems - causes and remedies

Problem

Causes

Remedies

Frost on chamber walls and mi-

crotome

- Cryostat is exposed to draft (open

windows, doors, air-condition-

ing).

- Sliding window has been open for

a long period of time at a very low

chamber temperature.

- Frost build-up caused by breath-

ing into the cryochamber.

- Eliminate draft sources or

change place of installation

of the cryostat.

- Close sliding window.

- Consider wearing a mask.

Sections smear

- Specimen not cold enough.

- Knife and/or anti-roll plate not yet

cold enough - sections melt.

- Select lower temperature.

- Wait until knife and/or

anti-roll plate have reached

chamber temperature.

Sections splinter

Sections not properly flattened

- Specimen too cold.

- Static electricity / draft.

- Specimen not cold enough.

- Large surface specimen.

- Anti-roll plate poorly adjusted.

- Anti-roll plate poorly aligned rela-

tive to knife edge.

- Wrong clearance angle

selected.

- Knife blunt or damaged.

- Select higher temperature.

- Eliminate cause.

- Select lower temperature.

- Trim specimen parallel; in-

crease section thickness.

- Readjust anti-roll plate.

- Align correctly.

- Set correct clearance angle.

- Use different part of knife or

replace knife.

Sections not properly flattened

despite correct temperature and

correctly aligned anti-roll plate

- Knife and/or anti-roll plate dirty.

- Edge of anti-roll plate damaged.

- Blunt knife.

- Clean with dry cloth or brush.

- Replace anti-roll plate.

- Use different part of knife or

replace knife.