Leica CM3050 S – Cryostat

6. Troubleshooting, applications tips

Problem

Causes

Remedies

Sections curl on anti-roll plate.

- Anti-roll plate does not protrude

far enough beyond the knife

edge.

- Readjust correctly.

Scraping noise during section-

ing stroke and specimen return

stroke.

- Anti-roll plate protrudes too far

beyond the knife edge and is

scraping against the specimen.

- Readjust correctly.

Ondulated sections.

- Knife damaged.

- Edge of anti-roll plate damaged.

- Use different part of the knife

or replace knife.

- Replace anti-roll plate.

Chatter marks form during

sectioning.

- Specimen insufficiently frozen

onto specimen disc.

- Specimen disc not clamped

tightly enough.

- Specimen holder ball joint not

clamped.

- Knife not clamped tightly enough.

- Selected section thickness too

thick - specimen has come off the

specimen disc.

- Specimen very hard and inhomo-

geneous.

- Blunt knife.

- Knife profile inappropriate for the

type of specimen to be sectioned.

- Wrong clearance angle.

- Refreeze specimen onto disc.

- Check disc clamping.

- Check ball joint clamping.

- Check knife clamping.

- Refreeze specimen onto disc.

- Increase section thickness;

if possible, reduce specimen

surface.

- Use different part of knife or

replace knife.

- Use knife with a different

profile.

- Readjust clearance angle

setting.

Condensation forms on anti-roll

plate and knife during cleaning.

- Brush, forceps, cloth or other

cleaning item too warm.

- Accessories and clean-

ing material to be used

pre-cooled only. - Store on

storage shelves inside the

cryochamber.