Instructions for Use V 1.7 RevC - 08/2014

5. Operation

2.

Cryochamber temperature:

Press key until menu item ‘Set chamber

temperature’ is displayed:
--> Continue as described for

specimen temperature.

3.

Preset counter

(instruments with sectioning motor only):

Press key until menu item ‘Preset coun-

ter’ is displayed:

= Preset counter activated

= Preset counter deactivated

Preset counter function:
The preset counter allows the user to carry out

a predefined number of sections in the ‘Contin-

uous stroke’ sectioning mode.

Once the preselected number of sections has

been carried out (= section counter reading

‘0’), the specimen head will stop automatically

in the lower limit position.

The reference value (i.e. the value from which

the preset counter counts downward) can be

selected in the corresponding menu item.

In order to be able to set a reference value, the

preset counter function must be active.

If the preset counter is deactivated (see dia-

grams above), press

to activate.

Now there are two options:
1. Increase or decrease the actual reference

value by pressing

or

.

2. Press and

simultaneously in order to

reload the former reference value (‘0’ when

setting up the instrument) and then press ei-
ther

or

individually in order to select

the desired reference value. Depending on

where you start from and where you want to

go, method 1 or method 2 is quicker (see ex-

amples).

Example 1:
Actual preset counter reading = 50

Former reference value = 300

New reference value of 65 to be selected.
Method 1 is most convenient:
• Increase from 50 to 65 by pressing

.

Example 2:
Actual preset counter reading = 50

Former reference value = 300

New reference value of 320 to be selected.
Method 2 is most convenient:
• Press

and

simultaneously to reload

old reference value (300):

• Increase from 300 to 320 by pressing

.

For further information on the preset counter,

please refer to chapter 5.3.6, Item 4 ‘Preset

counter’.

P R E S E T C O U N T E R
d n - - -

P R E S E T C O U N T E R
d n 0 5 6